Sample preparation

With such small sample sizes, care must be taken to ensure that collected samples are representative of the bulk material. This is especially relevant in rocks and minerals, which can vary widely in elemental content from region to region. Random, composite, and integrated sampling are each different approaches for obtaining representative samples.

Because ICP-MS can detect elements in concentrations as minute as a few nanograms per liter (parts per trillion), contamination is a very serious issue associated with collecting and storing samples prior to measurements. In general, use of glassware should be minimized, due to leaching impurities from the glass or absorption of analyte by the glass. If glass is used, it should be washed periodically with a strong oxidizing agent, such as chromic acid (H 2 Cr 2 O 7 ), or a commercial glass detergent. In terms of sample containers, plastic is usually better than glass, polytetrafluoroethylene (PTFE) and Teflon ® being regarded as the cleanest plastics. However, even these materials can contain leachable contaminants, such as phosphorus or barium compounds. All containers, pipettes, pipette tips, and the like should be soaked in 1 - 2% HNO 3 . Nitric acid is preferred over HCl, which can ionize in the plasma to form 35 Cl 16 O + and 40 Ar 35 Cl + , which have the same mass-to-charge ratios as 51 V + and 75 As + , respectively. If possible, samples should be prepared as close as possible to the ICP-MS instrument without being in the same room.

With the exception of solid samples analyzed by laser ablation ICP-MS, samples must be in liquid or solution form. Solids are ground into a fine powder with a mortar and pestle and passed through a mesh sieve. Often the first sample is discarded to prevent contamination from the mortar or sieve. Powders are then digested with ultrapure concentrated acids or oxidizing agents, like chloric acid (HClO 3 ), and diluted to the correct order of magnitude with 1 - 2% trace metal grade nitric acid.

Once in liquid or solution form, the samples must be diluted with 1 - 2% ultrapure HNO 3 to a low concentration to produce a signal intensity lower than about 10 6 counts. Not all elements have the same concentration to intensity correlation; therefore, it is safer to test unfamiliar samples on ICP-AES first. Once properly diluted, the sample should be filtered through a 0.25 - 0.45 μm membrane to remove particulates.

Gaseous samples can also be analyzed by direct injection into the instrument. Alternatively, gas chromatography equipment can be coupled to an ICP-MS machine for separation of multiple gases prior to sample introduction.

Standards

Multi- and single-element standards can be purchased commercially, and must be diluted further with 1 - 2% nitric acid to prepare different concentrations for the instrument to create a calibration curve, which will be read by the computer software to determine the unknown concentration of the sample. There should be several standards, encompassing the expected concentration of the sample. Completely unknown samples should be tested on less sensitive instruments, such as ICP-AES or EDXRF (energy dispersive X-ray fluorescence), before ICP-MS.

Questions & Answers

if three forces F1.f2 .f3 act at a point on a Cartesian plane in the daigram .....so if the question says write down the x and y components ..... I really don't understand
Syamthanda Reply
hey , can you please explain oxidation reaction & redox ?
Boitumelo Reply
hey , can you please explain oxidation reaction and redox ?
Boitumelo
for grade 12 or grade 11?
Sibulele
the value of V1 and V2
Tumelo Reply
advantages of electrons in a circuit
Rethabile Reply
we're do you find electromagnetism past papers
Ntombifuthi
what a normal force
Tholulwazi Reply
it is the force or component of the force that the surface exert on an object incontact with it and which acts perpendicular to the surface
Sihle
what is physics?
Petrus Reply
what is the half reaction of Potassium and chlorine
Anna Reply
how to calculate coefficient of static friction
Lisa Reply
how to calculate static friction
Lisa
How to calculate a current
Tumelo
how to calculate the magnitude of horizontal component of the applied force
Mogano
How to calculate force
Monambi
a structure of a thermocouple used to measure inner temperature
Anna Reply
a fixed gas of a mass is held at standard pressure temperature of 15 degrees Celsius .Calculate the temperature of the gas in Celsius if the pressure is changed to 2×10 to the power 4
Amahle Reply
How is energy being used in bonding?
Raymond Reply
what is acceleration
Syamthanda Reply
a rate of change in velocity of an object whith respect to time
Khuthadzo
how can we find the moment of torque of a circular object
Kidist
Acceleration is a rate of change in velocity.
Justice
t =r×f
Khuthadzo
how to calculate tension by substitution
Precious Reply
hi
Shongi
hi
Leago
use fnet method. how many obects are being calculated ?
Khuthadzo
khuthadzo hii
Hulisani
how to calculate acceleration and tension force
Lungile Reply
you use Fnet equals ma , newtoms second law formula
Masego
please help me with vectors in two dimensions
Mulaudzi Reply
how to calculate normal force
Mulaudzi
Got questions? Join the online conversation and get instant answers!
Jobilize.com Reply

Get Jobilize Job Search Mobile App in your pocket Now!

Get it on Google Play Download on the App Store Now




Source:  OpenStax, Physical methods in chemistry and nano science. OpenStax CNX. May 05, 2015 Download for free at http://legacy.cnx.org/content/col10699/1.21
Google Play and the Google Play logo are trademarks of Google Inc.

Notification Switch

Would you like to follow the 'Physical methods in chemistry and nano science' conversation and receive update notifications?

Ask